SAM V4.5 head phantoms has been designed for assessment of radiation patterns or total radiated power. Its geometry complies to the SAM data as defined by IEEE SCC 34, on which all relevant radiation standards are based.
|
| Construction | High precision injection molded PP Integrated positioning lines Sealed top cap for use with 17mm hexagonal key Compatible with all tissue simulating liquids including agressive solvents, e.g. DGBE type |
| Shell thickness | 2 ± 0.2mm (6mm at ear point) |
| Filling Volume | 7.2 liters |
| Weight | 1.2 kg (approx. 10kg filled with B900) |
| Dimensions | Width: 250mm (310mm with phone positioner) Depth: 269mm Height: 397mm incl. feet (Liquid height: 339 mm) |
| Phone Positioner | Integrated lightweight holder, accurate and easy to operate Enables precise positioning in both touch and 15° "cheek" position |
| Cap (black) | Resistant to all current tissue simulating liquids including DGBE recipes Option: Transparent cap for good control and easy elimination of air bubbles but not DGBE compatible |
| Mounting on Turntable | 8mm holes in a plane 210mm wide x 174mm deep |
| Compatibilities | The geometry complies with the SAM data as defined by IEEE SCC 34 3GPP TR25.914 “Measurements of Radio Performances for UMTS Terminals in Speech Mode” |
| Adapters | For EST OTA system now available |
| Options | Hand fixture/positioner Travel Case / Car Torso (TCCT) |