EASY4/MRI
easy4visual.jpg
EX3DV4 - Isotropic Dos-Probe

EX3DV4
Smallest Isotropic E-Field Probe for Dosimetric Measurements (Preliminary Specs)


Symmetrical design with triangular core
Built-in shielding against static charges
PEEK enclosure material (resistant to organic solvents, e.g., DGBE)
Calibration
ISO/IEC 17025 calibration service available.
Frequency 10 MHz to > 6 GHz
Linearity: ± 0.2 dB (30 MHz to 6 GHz)
Directivity ± 0.3 dB in HSL (rotation around probe axis)
± 0.5 dB in tissue material (rotation normal to probe axis)
Dynamic Range 10 µW/g to > 100 mW/g
Linearity: ± 0.2 dB (noise: typically < 1 µW/g)
Dimensions Overall length: 337 mm (Tip: 20 mm)
Tip diameter: 2.5 mm (Body: 12 mm)
Typical distance from probe tip to dipole centers: 1 mm
Application High precision dosimetric measurements in any exposure scenario (e.g., very strong gradient fields). Only probe which enables compliance testing for frequencies up to 6 GHz with precision of better 30%.
Compatibility DASY3, DASY4, DASY52 SAR and higher, EASY4/MRI

 

created and designed by R.Ø.S.A.