DASY52
probesvisual.jpg
ES3DV3 - Isotropic Dos-Probe

ES3DV3
Isotropic E-Field Probe for Dosimetric Measurements


Symmetrical design with triangular core
Interleaved sensors
Built-in shielding against static charges
PEEK enclosure material (resistant to organic solvents, e.g., DGBE)
Calibration
ISO/IEC 17025 calibration service available.
Frequency 10 MHz to 4 GHz;
Linearity: ± 0.2 dB (30 MHz to 4 GHz)
Directivity ± 0.2 dB in HSL (rotation around probe axis)
± 0.3 dB in tissue material (rotation normal to probe axis)
Dynamic Range 5 µW/g to > 100 mW/g;
Linearity: ± 0.2 dB
Dimensions Overall length: 337 mm (Tip: 20 mm)
Tip diameter: 3.9 mm (Body: 12 mm)
Distance from probe tip to dipole centers: 2.0 mm
Application General dosimetry up to 4 GHz
Dosimetry in strong gradient fields
Compliance tests of mobile phones
Compatibility DASY3, DASY4, DASY52 SAR and higher, EASY4/MRI

 

created and designed by R.Ø.S.A.