DASY52
dasyvisual.jpg
HAC RF Extension

HAC RF Extension for DASY5 AIR, PRO and NEO

Hearing Aid Compatibility (HAC) RF Extension: RF Emission Test of Wireless Devices (E- and H-field) according to ANSI-C63.19
Available for DASY5 AIR, PRO, NEO and DASY4

Contains the necessary hardware to test wireless devices for Hearing Aid Compatibility. Results are displayed in M-class levels as required for ANSI-C63.19.
Test Arch Phantom
Enables easy and well defined positioning of the phone and validation dipoles as well as simple teaching of the robot.
Dimensions: 370 x 370 x 370 mm
Phone Positioner Supports accurate and reliable positioning of any phone.
Effect on near field <± 0.5 dB

RF Emission Calibration Dipoles

CD700V3 - CD2450V3
The dipoles can be used to test hearing aids as described in ANSI-C63.19.
Product ISO/IEC 17025 calibration service available. 
Frequency Band

800-960 MHz; 1710-2000 MHz; 2250-2650 MHz (in free space)

Return Loss > 15 18 dB over frequency band
> 20 dB at calibration frequency
Options Dipoles for other frequencies or solutions and other calibration conditions upon request

Software

Easy teaching of predefined ANSI-C63.19 measurement area
Predefined measurement procedures for compliance and validation
Evaluation incorporates automatic exclusion of high-level areas
Documentation ready for inclusion into compliance report
Probes Isotropic free field H3DV6 H Probe and and ER3DV6 E Probe needed for RF field measurements. ISO/IEC 17025 calibration service available.