Contains the necessary hardware to test wireless devices for Hearing Aid Compatibility. Results are displayed in M-class levels as required for ANSI-C63.19. | |||
| Test Arch Phantom |
Enables easy and well defined positioning of the phone and validation dipoles as well as simple teaching of the robot. Dimensions: 370 x 370 x 370 mm |
|||
| Phone Positioner | Supports accurate and reliable positioning of any phone. Effect on near field <± 0.5 dB |
|||
|
RF Emission Calibration Dipoles |
CD700V3 - CD2450V3 The dipoles can be used to test hearing aids as described in ANSI-C63.19. |
|||
| Product | ISO/IEC 17025 calibration service available. |
|||
| Frequency Band |
800-960 MHz; 1710-2000 MHz; 2250-2650 MHz (in free space) |
|||
| Return Loss | > 15 18 dB over frequency band > 20 dB at calibration frequency |
|||
| Options | Dipoles for other frequencies or solutions and other calibration conditions upon request | |||
|
Software |
Easy teaching of predefined ANSI-C63.19 measurement area Predefined measurement procedures for compliance and validation Evaluation incorporates automatic exclusion of high-level areas Documentation ready for inclusion into compliance report |
|||
| Probes | Isotropic free field H3DV6 H Probe and and ER3DV6 E Probe needed for RF field measurements. ISO/IEC 17025 calibration service available. | |||