Measurement » Probes » H3DV7 - Isotropic H-Probe

H3DV7
3-Dimensional H-Field Probe for MRI and other Low Frequency Applications

Download Product Flyer (PDF, 583 kB)

Construction Three concentric loop sensors with 3.8 mm loop diameters (3 turns each)
Resistively loaded detector diodes for linear response
Built-in shielding against static charges
PEEK enclosure material (resistant to organic solvents, e.g., glycolether)
Frequency 40 MHz to 1 GHz (absolute accuracy ± 6.0%, k=2); Output linearized (can be extended to < 20 MHz)
Directivity < ± 0.2 dB (spherical isotropy error); < ± 0.4 dB for f < 100 MHz
Dynamic Range 0.01 A/m to 10 A/m at 128 MHz
E-Field Interference < 10% at 1 GHz (for plane wave)
Dimensions Overall length: 330 mm (Tip: 40 mm)
Tip diameter: 6 mm (Body: 12 mm)
Distance from probe tip to dipole centers: 3 mm
Application General magnetic near-field measurements up to 1 GHz (in air or liquids)
Field component measurements
Measurement of RF fields within MRI
Surface current measurements
Low interaction with the measured field


 
© 2008 by SPEAG, Schmid & Partner Engineering AG, CH-8004 Zurich, Switzerland | SitemapContact | Login