Measurement »
Probes »
H3DV6 - Isotropic H-Probe
H3DV6
3-Dimensional H-Field Probe for Small Band Applications
Download Product Flyer (PDF, 583 kB)
| Construction | Three concentric loop sensors with 3.8 mm loop diameters Resistively loaded detector diodes for linear response Built-in shielding against static charges PEEK enclosure material (resistant to organic solvents, e.g., glycolether) |
| Frequency | 200 MHz to 3 GHz (absolute accuracy ± 6.0%, k=2); Output linearized |
| Directivity | ± 0.2 dB (spherical isotropy error) |
| Dynamic Range | 10 mA/m to 2 A/m at 1 GHz |
| E-Field Interference | < 10% at 3 GHz (for plane wave) |
| Dimensions | Overall length: 330 mm (Tip: 40 mm) Tip diameter: 6 mm (Body: 12 mm) Distance from probe tip to dipole centers: 3 mm |
| Application | General magnetic near-field measurements up to 3 GHz (in air or liquids) Field component measurements Surface current measurements Low interaction with the measured field |