Measurement »
Probes »
ET1DV4 - Dos-Probe
ET1DV4
Single-Sensor E-Field Probe for Isotropic Dosimetric Measurements
Download Product Flyer (PDF, 583 kB)
| Construction | Single sensor with 35.3º rotated directivity Sequential measurement of the three components Built-in shielding against static charges PEEK tip |
| Calibration | Basic Broad Band Calibration in air Conversion Factors (CF) for HSL 900 and HSL 1810 Additional CF for other liquids and frequencies upon request |
| Frequency | 10 MHz to 10 GHz Linearity: ± 0.2 dB (100 MHz to 10 GHz) |
| Directivity | ± 0.1 dB in HSL (rotation around probe axis) ± 0.4 dB in HSL (rotation normal to probe axis) |
| Dynamic Range | 20 µW/g to > 100 mW/g; Linearity: ± 0.2 dB |
| Dimensions | Overall length: 330 mm (Tip: 12 mm) Tip diameter: 1.2 mm (Body: 12 mm) Distance from probe tip to dipole centers: 0.8 mm |
| Application | Dosimetry in cell dishes or small animals Dosimetry up to 10 GHz in very strong gradient fields Dosimetry near material boundaries Very low interaction with the measured field Needs positioner which can do accurate probe rotations |